background image

7

Measuring Resistance

• 

Insert the red test lead into the Ω terminal and the black test lead into the COM 

terminal.

• 

Set the rotary switch to Ω resistance measurement mode.

• 

Connect the test leads across with the object being measured. The measured 

value shows on the display.

• 

For high resistance measurement >1MΩ, it is normal for it to take several seconds 

to obtain a stable reading.

Warning

 

To avoid damages to the Meter or to the devices under test, disconnect circuit power 

and discharge all the high-voltage capacitors before measuring resistance.

To avoid harm to yourself, do not input higher than DC 60V or AC 30V voltages.

Notes

 

The display will show “1” if the tested resistance is greater than the maximum range of 

the meter.

• 

When resistance measurement has been completed, disconnect the connection 

between the testing leads and the circuit under test, and remove the testing leads 

away from the input terminals of the Meter.

Capacitance measurement

• 

Set the rotary switch to F to select capacitance measurement.

• 

Insert the tested capacitor directly into the capacitance jacks or use test leads.

• 

Inserting the capacitor directly is more stable and suitable for small value of 

capacitance testing.

• 

“1” is displayed if the tested capacitor is shorted or exceeds the maximum range.

• 

For high capacitance measurement, it is normal for it to take several seconds to 

obtain a stable reading.

• 

When capacitance testing has been completed, disconnect the connection 

between the testing leads and the item under test, and remove the testing leads 

away from the input terminals of the Meter.

Warning

 

To avoid damages to the Meter discharge all the high-voltage capacitors before 

measuring.

Testing Diodes and Continuity

• 

Insert the red test lead into the 

 terminal and the black test lead into the COM 

terminal.

• 

Set the rotary switch to 

 to select diode and/or continuity test mode.

• 

For forward voltage drop readings on any semiconductor component place the red 

test lead on the anode and the black test lead onto the cathode.

• 

The measured value shows on the display, or if measuring continuity and the 

resistance is below 30Ω the internal buzzer will sound.

• 

When testing has been completed, disconnect the connection between the testing 

leads and the item under test, and remove the test leads from the input terminals 

of the Meter.

Note

: A good diode should produce a forward voltage drop of 0.5V to 0.8V.

Summary of Contents for 72-2600

Page 1: ...1 Model 72 2600 Digital Multimeter ...

Page 2: ... shorter than 10 seconds For current values over 5A the wait period between each measurement must be longer than 15 minutes When a measurement has been completed disconnect the testing probes from the circuit under test Replace the batteries as soon as the low battery indicator appears on the display Remove dead batteries from the meter or if it is not going to be used for a long time Never mix ol...

Page 3: ...cy Input Impedance Overload protection 200mV 10µV 0 05 3 Around 10MΩ 600VAC DC RMS 850V p p 2V 100µV 0 1 3 20V 1mV 0 15 5 200V 10mV 600V 10mV 1 2 3 4 5 AC VOLTAGE Range Resolution Accuracy Input Impedance Overload protection 200mV 10µV 0 8 10 Around 10MΩ 600VAC DC RMS 850V p p 2V 100µV 0 5 10 20V 1mV 0 6 10 200V 10mV 600V 100mV 0 8 15 Displays effective value of sine wave AC voltage frequency rang...

Page 4: ...ent is allowed When 5A use 10 seconds measurement at an interval of more than 15 minutes Measured voltage drop 200mV for full range AC current frequency range 40hz 400hz Display true RMS value waveform coefficient is not more than 5 RESISTANCE Range Resolution Accuracy Overload Protection 200Ω 0 01Ω 0 5 10 600V AC DC RMS 2kΩ 0 1Ω 0 3 3 20kΩ 1Ω 0 3 1 200kΩ 10Ω 2MΩ 100Ω 20MΩ 1kΩ 0 5 1 Open circuit v...

Page 5: ...cuit voltage approx 3 0V OPERATION Rotary Switch Below table indicates the rotary selector switch position functions DC Voltage measurement AC Voltage measurement AC Current measurement DC Current measurement F Capacitance measurement Hz Frequency measurement Diode Test Continuity sounder Resistance measurement Ω A A Hold Button Press to enter or exit data hold mode The Hold mode is applicable to ...

Page 6: ...reduce it step by step until a satisfactory reading is obtained Do not attempt to measure higher than 600V or the meter may be damaged and there may be a risk of electric shock Disconnect from the circuit under test when changing the range If the display only shows 1 it indicates the measured voltage is out of range In this case turn the range selector to a higher range Measuring DC Current Insert...

Page 7: ...leads Inserting the capacitor directly is more stable and suitable for small value of capacitance testing 1 is displayed if the tested capacitor is shorted or exceeds the maximum range For high capacitance measurement it is normal for it to take several seconds to obtain a stable reading When capacitance testing has been completed disconnect the connection between the testing leads and the item un...

Page 8: ...ter CLEANING Frequency measurement Insert the red test lead into the Hz terminal and the black test lead into the COM terminal Set the rotary switch to 20kHz frequency measurement Connect the test leads across the object being measured and the value shows on the display When frequency measurement has been completed disconnect the connection between the testing leads and the circuit under test and ...

Reviews: