Tenma 72-6851 & 72-6853 Instruction manual
Power on Settings
Most instrument settings are stored in non-volatile ram and will remain unchanged while power is
off. The following instrument status values are set at power on:
Status Byte Register
= 0
*
Service Request Enable Register
= 0
Standard Event Status Register
= 128 (pon bit set)
*
Standard Event Status Enable Register
= 0
Limit Event Status Register
= 0 (Then set to show new limit status)
*
Limit Event Status Enable Register
= 0
Execution Error Register
= 0
Query Error Register
= 0
*
Parallel Poll Enable Register
= 0
*
Registers marked thus are specific to the GPIB section of the instrument and are of limited
use in an ARC environment.
The standby state will be set and the instrument will be in local with the keyboard active. A
complete self-test will be performed just after power up and this will determine the values in the
Failed Output Register and the state of the FLT bit in the Status Byte Register.
Due to the non-volatile storage the power on settings are essentially modified by any command,
local or remote, which changes any value not listed above. If a defined state is require at start up
by the controller the
*
RST command should be issued which will load the settings as listed in the
description of that command.
If for any reason an error is detected at power up in the non-volatile ram a warning will be issued
and all settings will be returned to their default states as for a
*
RST command.