To replace the Meter’s fuse:
1. Turn the Meter off and remove all connections from
the terminals.
2. Remove the holster from the Meter.
3. Remove the 3 screws from the case bottom, and
separate the case top from the case bottom.
4. Remove the fuse by gently prying one end loose,
then take out the fuse from its bracket.
5. Install ONLY replacement fuses with the identical type
and specification and make sure the fuse is fixed
firmly in the bracket.
Fuse: 0.5A, 250V, fast type,
φ
5x20mm
6. Rejoin the case bottom and case top, and reinstall
the 3 screws and holster.
An open fuse is always the result of improper operation.
Maintenance(2)
C.Replacing the Battery
(see figure 10)
To replace the Meter’s battery:
1. Turn the Meter power off and remove all connections
from the terminals.
2. Remove the holster from the Meter.
3. Remove the 3 screws from the case bottom, and
separate the case top from the case bottom.
4. Remove the battery from the battery connector.
5. Replace with a new 9V battery (NEDA1604, 6F22
or 006P).
6. Rejoin the case bottom and case top, and reinstall
the 3 screws and the holster.
Warning
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery
as soon as the battery indicator “ ” appears.
Model 72-7930: OPERATING MANUAL
33
Summary of Contents for 72-7930
Page 9: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 21: ...Model 72 7930 OPERATING MANUAL 21 To avoid personal injury F ...
Page 23: ...Model 72 7930 OPERATING MANUAL 23 F hFE To avoid personal injury and mA ...
Page 24: ...POWER Model 72 7930 OPERATING MANUAL 24 100 ...
Page 25: ...Model 72 7930 OPERATING MANUAL 25 ...
Page 29: ...Model 72 7930 OPERATING MANUAL 29 100 ...