Quick-Start Operating Guide 

 

 

High- Ionizing Modules for Terra Desiccators 

© 

Copyright 2010 Terra Universal Inc.  All rights reserved. • Revised Sept. 2010 • Document No. 1800-39

v2.0tm

 

Terra Universal, Inc.  •  TerraUniversal.com  •  800 S. Raymond Ave.  •  Fullerton, CA  92831  •  TEL: (714) 578-6000  •  FAX: (714) 578-6020 

 

 

 

 

 

 

1.0

 

Introduction 

 

This manual provides information 

on installing and operating your 

Terra Universal Ionizing Module. 
 

By studying this document 

carefully, you can be assured of a 

long, efficient service life from 
your system.

 

 

 

2.0

 

Description 

 

Terra’s Desiccator Ionizing Module incorporates pairs of ion emitters in each desiccator chamber that generate a 

balanced stream of positive and negative ions to safely neutralize static charges on surfaces sensitive to electro-static 
discharge. This module is designed for installation in a pressurized (purged) cabinet. The purge gas distributes the 

positive and negative up to 48" from the emitter tips (see “Specifications” for static decay rates). 

 

Because the Ionizing Module operates at low voltage (24 VAC), it is safe to operate and generates minimal RFI/EMI. 
The stainless steel case contains the low-voltage power supply and titanium emitters, which can be replaced in a 

matter of minutes.

 

 

 

Note:

  Refer to QuickStart document No. 1800-45 for ionizing equipment used for hoods, modular cleanrooms, and 

glove boxes. 

 

3.0

 

Installation 

 

Carefully unpack your desiccator cabinet (refer to separate QuickStart for desiccator set-up). Report any shipping 

damage immediately to the freight company. 

 

The Ionization Module is installed in the desiccator plenum chamber before shipping. Use the low-voltage connector 
and power cable included to connect the Module to the power transformer. Then connect the transformer to the 

appropriate power supply (120VAC or 240VAC). A single transformer provides ample power for up to 20 emitter pairs. 

Proprietary Notice

 

 

This manual pertains to proprietary devices manufactured by Terra 
Universal, Inc. Neither this document nor any portion of it may be 
reproduced in any way without prior written permission from Terra 
Universal.

 

Terra Universal makes no warranties applying to information contained 
in this manual or its suitability for any implied or inferred purpose.  
Terra Universal shall not be held liable for any errors this manual 
contains or for any damages that result from its use.

 

Safety Notice

 

 

CAUTION 

Cautions are used when failure to observe instructions 
could result in significant damage to equipment. 

 

 

A thorough familiarity with all operating guidelines is essential to safe 
operation of the product.  Failure to observe safety precautions could 
result in poor performance, damage to the system or other property, or 
serious bodily injury or death. 
 
The following symbols are intended to call your attention to two levels 
of hazard involved in operation: 

WARNING 

Warnings are used when failure to observe instructions 
or precautions could result in injury or death. 

The information presented here is subject to change without notice. 

Plenum holes 
for gas 
distribution 

Summary of Contents for 2006-01

Page 1: ... No 1800 39 Ionizing Modules for Terra Desiccators Copyright 2010 Terra Universal Inc All rights reserved Revised September 2010 Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 ...

Page 2: ...uickStart for desiccator set up Report any shipping damage immediately to the freight company The Ionization Module is installed in the desiccator plenum chamber before shipping Use the low voltage connector and power cable included to connect the Module to the power transformer Then connect the transformer to the appropriate power supply 120VAC or 240VAC A single transformer provides ample power ...

Page 3: ...ciency In this case replace the tips Use a pencil eraser or needle nosed pliers to remove the old tip Insert the new tip and gently press until it seats against the connector no twisting is necessary Cat Replacement Titanium Emitters 2005 45 The emitter housing may be cleaned using alcohol or other mild detergent and de ionized water 5 0Specifications Emitter Points Replaceable machined titanium 3...

Page 4: ...aid for separately For Terra s Policy and Procedures for Returning Goods see Terra s Internet site www TerraUniversal com Generally customers can improve the chance of collecting on a freight claim by following these procedures 1 formally requesting that the carrier inspect the shipment immediately upon suspecting damage or shortage to verify condition 2 notifying the carrier upon discovery of con...

Reviews:

Related manuals for 2006-01