Installation on Laminar Flow Hoods 

 

When ordered with a laminar flow hood, the IonBar™ is installed in the exact same way as with a fan/filter unit in a 

cleanroom.  The mounting brackets will rest on the top edges of the hood and the FFU will be placed directly on top of 

the brackets, holding the IonBar™ in position.  The IonBar™ emitters will aim downwards toward the work surface. 

 

 

 

Whether incorporated into a laminar-flow system or used as a stand-alone device, the IonBar™ requires connection 

to a 24V power transformer.  If the IonBar™ is ordered with the Installation Kit, the power transformer, power cable, 

and appropriate mounting clips will be included. 

 

1.

 

Use a telephone-type connector and low-voltage cable to connect the IonBar™ to the power transformer.  

2.

 

Connect the transformer to the appropriate power supply (120VAC or 240VAC).  

 

 

NOTE

 

A single transformer provides ample power for up to 20 emitter pairs. Depending on the length 

of your IonBar™, multiple units can be interconnected using low-voltage telephone cable to 

daisy-chain the units. 

 

Once connected to power, the IonBar™ immediately emits balanced streams of positive and negative ions through 

each emitter pair. Effective ion distribution requires an airflow speed of at least 50 fpm. 

 

 

Do not block access to the power transformer.  This connection serves as the main disconnect 

for the system in the event of an emergency. 

CAUTION 

Figure 3: 

Place the IonBar™ on the hood frame before lowering the fan/filter unit into position. 

Summary of Contents for IonBar

Page 1: ...ide Document No 1800 45 IonBar Ionizing Bar Copyright 2016 Terra Universal Inc All rights reserved Revised November 2016 Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 ...

Page 2: ...ods Preinstalled IonBars can also be purchased individually and installed by the user in the desired application An installation kit with standard FFU mounting brackets is sold separately and includes the power supply transformer NOTE For installation in a hood the overall hood width should be at least 2 greater than the length of the desired IonBar Proprietary Notice This manual pertains to propr...

Page 3: ...e mounting brackets installed place the IonBar in the ceiling bay where the fan filter unit goes The mounting brackets will rest upon the frame of the ceiling bay no other fasteners are required Simply lower the FFU into the ceiling bay and allow the weight of the FFU to hold the IonBar in position Installation on Gloveboxes When ordered with a glovebox the IonBar will be mounted to a stainless st...

Page 4: ...r transformer If the IonBar is ordered with the Installation Kit the power transformer power cable and appropriate mounting clips will be included 1 Use a telephone type connector and low voltage cable to connect the IonBar to the power transformer 2 Connect the transformer to the appropriate power supply 120VAC or 240VAC NOTE A single transformer provides ample power for up to 20 emitter pairs De...

Page 5: ...emove deposits from the emitter points 3 If deposits cannot be removed or if the tips appear worn or bent the IonBar may not discharge surfaces with optimal efficiency In this case replace the tips Use a pencil eraser or needle nosed pliers to remove the old tip Insert the new tip and gently press until it seats against the connector no twisting is necessary 4 0 Specifications Replacement Parts Pr...

Page 6: ...0Hz Power Rating 1A List of Input Connections Telephone cable connector low voltage List of Output Connections Telephone cable connector wall transformer Equipment Installation Assembly See Section 2 Installation Page 3 Protective Earthing Grounding N A Utility Requirements None Equipment Operation Operating Controls None Disconnect Device Wall transformer see warning on Page 5 Accessory Connectio...

Page 7: ...separately For Terra s Policy and Procedures for Returning Goods see Terra s Internet site www TerraUniversal com Generally customers can improve the chance of collecting on a freight claim by following these procedures 1 formally requesting that the carrier inspect the shipment immediately upon suspecting damage or shortage to verify condition 2 notifying the carrier upon discovery of concealed d...