PROBLEM
POSSIBLE CAUSES
POSSIBLE SOLUTION
Intermittent output
Lead wire
Verify connection is secure. Insert wire firmly.
Turn down the intensity. Rotate lead wire in socket 90°. If still intermittent,
replace lead wire.
If still intermittent after replacing lead wire, a component may have failed. Call
the repair department.
Stimulation is ineffective.
Improper electrode placement.
Reposition electrode.
Unknown
Contact clinician.
The skin becomes
red and/or you feel
a stabbing pain.
Using electrodes on the same site every time.
Reposition the electrodes. If at any time you feel pain or discomfort, stop use
immediately.
Electrodes are not adhered to the skin properly.
Ensure the electrodes are securely adhered to the skin.
The electrodes are dirty.
Clean the electrodes according to description in this manual or replace with
new electrodes.
The surface of the electrode is scratched.
Replace with new electrodes.
Output current stops
during therapy.
The electrodes come off the skin.
Turn off the device and place the electrodes on again, or replace with new
electrodes.
The lead wire is disconnected.
Turn off the device and connect the lead wire.
The batteries’ power has been exhausted.
Replace with new batteries.