TABLE 3:
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of these electrical stimulators should assure that it is used in such environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
Not applicable
3 V/m
80 MHz to
2,5 GHz
3 V/m
Portable and mobile RF
Communications equipment should be used no closer to any part of the device,
including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1,2
d = 1,2 80 MHZ to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
Where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and
d
is the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol: