Page
46
of
76
Date
01-01-2010
–
31-12-2099 (DD-MM-YYYY)
Sets the date of the internal clock.
When T-RAC II is connected to TOMEs via the network,
the date is automatically taken from TOMEs.
Manage progs
Both / TOMs
Default: Both
Both:
Programs can be managed on both T-RAC II and TOMEs.
TOMs:
Programs can only be managed on TOMEs. On T-RAC II, it is not possible to
edit / duplicate / delete programs or to change the order of the programs.
This option provides greater certainty that all devices contain the same
programs.
The programs can still be loaded/saved via the USB flash drive.
TOMEs short
1
–
999
Default: 1
Only useful when using a TOMEs network. On TOMEs, T-RAC II devices are referenced
with their TOMEs short numbers instead of their serial numbers. It is a unique number
for each T-RAC II device in the network. When two devices with the same number get
connected, TOMEs will automatically give one of the devices a new TOMEs short number.
[NOTE] Changing this settings is disabled while T-RAC II is connected to TOMEs.
Discon
nect from TOMEs (e.g. by setting the program setting ‘Enable LAN’ to ‘No’) before
changing this setting.
7.2
Programs
T-RAC II Jr has one program which cannot be transferred to other devices.
T-RAC II Ethernet and Wifi can contain up to ten programs. The programs are
interchangeable with other devices and with TOMEs via the TOMEs network or via a USB
flash drive. At startup, T-RAC II will always load the first program in the program list
(program number 1). Switching to another program is possible when the even
t ‘Program
Summary of Contents for T-RAC II
Page 68: ......
Page 69: ......
Page 70: ... Nhãn hiệu đã đăng ký Bản quyền Terumo BCT Inc Tất cả các quyền được bảo hộ ...
Page 71: ...02 2020 ...
Page 72: ......
Page 144: ...Page 74 of 76 14 Symbol LS 2208 barcode reader set up ...
Page 145: ...Page 75 of 76 ...