PVFlow. delay
Default: 00:00
00:00
–
05:00 (mm:ss)
Sets the interval between venipuncture (moment of pressing ‘start’
button) and
the moment blood should flow into the bag. If this time is exceeded, an alarm will
sound. If set to 00:00, the post venipuncture flow delay alarm is disabled and no
audible low flow alarm will be given until the first blood is collected.
Go PVFlow. de
Yes / No
Default: Yes
Yes:
When the post venipuncture flow delay time is exceeded, the user can
still continue the donation.
No:
When the time set in PVFlow. delay is exceeded, the donation will be
cancelled automatically.
Clamp interval
Yes / No
Default: No
Yes:
After donation, the clamp will go to the ‘locked’ state for 10 seconds
each minute. In this way, more blood might be donated but the risk
of coagulation in the area of the venipuncture is less.
No:
After
donation, the clamp stays closed until the user presses ‘open
clamp’.
Time tubes
Yes / No
Default: No
Yes:
When collecting blood in tubes before donation, this is normally done
on the venipuncture screen. On that screen, an extra button will be
shown which should be pressed between venipuncture and the
collecting of the blood tubes. In this way, both actions can be timed.
These durations are recorded into the donation data.
Time tubes button
No:
No extra button or timer is activated.