Page
64
of
76
Bi-directional communication: info from BBIS
Indicated by:
Message on the touch screen.
Cause:
Via the BBIS, a program and/or a volume was set to be used
during the donation. The program and/or volume is depicted on
the screen.
Number and name of the selected program.
Collection volume set by the BBIS.
Solution:
Check whether the program and volume are OK. If so, continue
the ongoing donation by pressing
. When pressing
, the donation will automatically be cancelled.
Bi-directional communication: info from BBIS not plausible
Indicated by:
Message on the touch screen.
Cause:
Via the BBIS, a volume was set to be used during the donation.
The volume which was selected on the BBIS, is not within the
range allowed in the program on T-RAC II.
Solution:
The information from the BBIS needs to be resent (to be
performed on the BBIS). Otherwise, the donation will be
cancelled.
Bi-directional communication: donor approval not OK
Indicated by:
Message on the touch screen.
Cause:
On the BBIS, the doctor has refused the donor to donate.
Solution:
Press ‘Continue’, the donation will be cancelled automatically.
Summary of Contents for T-RAC II
Page 68: ......
Page 69: ......
Page 70: ... Nhãn hiệu đã đăng ký Bản quyền Terumo BCT Inc Tất cả các quyền được bảo hộ ...
Page 71: ...02 2020 ...
Page 72: ......
Page 144: ...Page 74 of 76 14 Symbol LS 2208 barcode reader set up ...
Page 145: ...Page 75 of 76 ...