Maintenance and Inspections
6.
While holding the lid of the Fast Clamp, pull up the lever
until you hear a click. (Figure 5-12)
• After pulling up the lever, the lever has returned to the original position.
• If the lever does not return to the original position, pull down the lever. If it does
not return to the original position, a failure may have occurred.
Stop using the system immediately, and contact Terumo’s service representative.
Figure 5-12
Inspection method of the battery
This system has a built-in battery as a back up power source.
As the battery deteriorates regardless of use, periodical replacement is necessary.
Replacement period of the battery differs based on the environment and frequency of use of the Controller. The battery should be
replaced if an abnormality is found in the following inspection. It should also be replaced after approximately two years of use, even
if no abnormality is found.
Inspect the battery once every six months with the following procedures:
1.
Connect the AC power cable to the Controller.
2.
Plug in the AC power cable to the grounded AC power
source.
3.
Charge the battery until the charge lamp turns off.
(Figure 5-13)
4. Turn
on
the
power.
Figure 5-13
• There is more than 70% battery charge level.
127
Check
Check
Note
Six-month Inspection
Fast Clamp
Lever
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...