Operation Procedure
3.
Close the lid of the clamp-on ABD/Flow sensor.
(Figure 1-37)
Press the lid until you hear a click.
Figure 1-37
This section describes the step for using the temperature measuring function. When using the temperature measuring function,
confirm that the setup of the temperature measuring function of the channel to be used is on. (Refer to “Confirming the Setup”
(page 48).)
1.
Attach the connector of the Temperature Sensor Cable to
the CAPIOX Luer Thermistor which is connected to the
extracorporeal circulation circuit line. (Figure 1-38)
Figure 1-38
This section describes the step for using the pressure measuring function. When using the pressure measuring function, confirm
that the setup of the pressure measuring function of the channel to be used is on. (Refer to “Confirming the Setup” (page 48).)
1. Attach the connector of the Pressure Sensor Cable to the
pressure sensor which is connected to the extracorporeal
circulation circuit line. (Figure 1-39)
Figure 1-39
1.
Fill the extracorporeal circulation circuit with the priming solution by gravity.
46
Attaching the Temperature Sensor (CAPIOX Luer Thermistor)
Attaching the Pressure Sensor
Filling Priming Solution
Lid
CAPIOX Luer Thermistor
Connector
Pressure sensor
Connector
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...