Other Operation Procedures
(This example describes when pump coast is activated when the Over Pressure Alarm is issued for P1.)
Setting of the Controller
1.
Turn the P1 pressure measuring function on.
For details about how to set, refer to “Setting Pressure Measuring Function”
(page 85).
2.
Set the safety connection of the P1 Over Pressure Alarm to
“Pump Coast”.
For details about how to set, refer to “Setting the Safety Connection”
(page 91).
• When activating the safety connection, be sure to also turn on the pressure measuring function. The safety connection is not
activated without the pressure measuring function.
Actions after activating the safety connection
1.
When the Pressure Alarm is issued, the motor speed is
lowered to the coast speed.
The status lamp starts to flash in red and the alarm sounds, while “P1:Over
Pressure Alarm” and “P1 Over: Pump Coasting” are displayed on the
message area. (Figure 2-1)
• When the motor speed is equal to or less than the coast speed, the safety
connection is not activated.
2.
Remove the cause of the Pressure Alarm.
For details about how to handle, refer to “Alarms and Troubleshooting”
(page 132).
Figure 2-1
3.
While monitoring the pressure and the flow, adjust the
motor speed to achieve the necessary flow.
The status lamp is lit in green, and the alarm sound stops.
“P1 Over: Pump Coasting” on the message area is turned off.
66
a
When the safety connection of the Pressure Alarm (over pressure/low pressure) is set to “Pump Coast”
Note
Note
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...