Why does the indication “ ” remain (unchanged)
although blood was applied to the Test TIP?
The blood sample
volume is not sufficient.
The blood does not reach the reagent
pad (white part) in the Test TIP.
A
Blood sample was too small.
Wipe off the blood from the puncture site and
attach a new Test TIP to start a completely new
measurement. After you have pricked the puncture
site, immediately draw up a sufficient amount of
blood.
Note well
The blood sample
volume is sufficient.
The blood spreads over the reagent pad
(white part) in the Test TIP.