Indication
Cause
Pag.
Test TIP is inserted at an angle (not
straight), or Test TIP is not fully
inserted.
Push the Test TIP firmly and correctly onto
the Meter before drawing up the blood.
18,19
When blood is on the Test TIP, replace
the Test TIP with a new one and take
measurement again.
Replace the Test TIP with a new one and take
measurement again.
When the below indications appear, measurement cannot be done in a proper way.
Please conduct the following to solve the problem.
Clean the measuring window with a cotton swab.
Replace the Test TIP with a new one and take
measurement again.
If “
” is still shown, even though the above processes are conducted, please contact TERUMO.
Remove the Test TIP and check the
measuring window.
If it is clean, the indication will disappear.
When the indication disappears, attach the
TIP and take measurement.
If it is dirty, the indication will not disappear.
Clean the measuring window with a cotton swab.
After cleaning, conduct the Meter operation check
process by using the Meter-Check TIP.
Used Test TIP is attached on the
meter.
Measuring window is dirty or
dusty.
Measuring window may be dirty
or dusty.
16,17
16,17
44
16,17
44
52,53
How to deal with it