TECHNICAL INFORMATION
For SMP2 serial number 6000 and greater
The Harvest SmartPReP 2 Centrifuge meets the requirements for leakage current and hi-pot testing listed
in BS EN 60601-1 and UL 2601-1 standards for Medical Electrical Equipment.
(Less than 100µA on model SMP2-100 and SMP2-115)
(Less than 500µA on model SMP2-230)
G force during first spin = 1250
G force during second spin = 1050
Disconnection device
Appliance coupler
Rotor Disposable Capacity
2 or 4
Processing Time
approx. 14 minutes
The appliance coupler (also known as the power cord) must always be easily accessible. 2.5 meter power
cords supplied by Harvest Technologies, Corp. are the only acceptable means to power the SmartPReP 2
Centrifuge. Power quality should be that of a typical commercial or hospital environment.
PHYSICAL SPECIFICATIONS
Height:
8.73 in. (22.2 cm)
Weight:
34 lb. (15.42 kg)
Width:
16.5 in. (41.91 cm)
Weight, Ship:
42 lb. (19.05 kg)
Depth:
18 in. (45.72 cm)
Harvest reserves the right to discontinue or change specifications without prior notice on all products. The
SmartPReP 2 System has been tested and conforms to the following standards:
STANDARDS
Standard #
Description
UL 61010-1
Issued:2004/07/12 Ed:2 Rev:2008/10/28 UL Standard for Safety Electrical
Equipment For Measurement, Control, and Laboratory Use; Part 1: General
Requirements
CAN/CSAC22.2#61010-1
Issued: 2004/07/01 Ed: 2 Standard for Safety Electrical Equipment For
Measurement, Control, and Laboratory Use; Part 1: General Requirements
Included when done concurrently with UL/EN/IEC 61010-1.
IEC 61010-2-020
Issue: 2006/05/01 Ed: 2 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part 2-020: Particular Requirements
for Laboratory Centrifuges; Amd. 1-1996
CENELEC EN 61010-1
Issued:1993/04/01 Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use - Part 1: General Requirements-
Includes Amendment A2:1995; IEC 61010-1:1990 + A1:1992, Mo A2:1995
Included when done concurrently with UL/CSA/IEC 61010-1
IEC 61010-1
Issued: 2001/02/01 Ed: 2 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use - Part 1: General Requirements;
Corrigendum 1:05/2002; Corrigendum 2:04/2003
IEC 60601-1-2
Issued:2007/03/01 Ed:3.0 Medical Electrical Equipment - Part 1-2: General
Requirements for Basic Safety and Essential Performance - Collateral Standard:
Electromagnetic Compatibility - Requirements and Tests