60
Changing Settings on the Menu
Dial
Increases value
Decreases
value
6)
Turn the Dial to set the bolus flow rate.
Note
• The bolus flow rate can be set by 100 mL/h in the following value range.
20 drop/mL: 100.00 to 300.00 mL/h
60 drop/mL: 100.00 mL/h (fixed)
Ready for bolus icon
Operation indicator (green flashing)
Purge switch
Bolus while being
pressed
7)
Press and hold the Purge switch to start the bolus.
Notes
• The operation indicator flashes green.
• The bolus volume delivered increases.
• The buzzer sounds during solution delivery.
Bolus volume delivered
Volume delivered
8)
Release the Purge switch to stop the bolus.
Notes
• If the Purge switch is pressed again within 1.5 seconds after being released,
Hands On Bolus continues and the bolus volume delivered display is maintained.
• If exceeding 1.5 seconds after the Purge switch has been released, the screen
returns to the flow rate screen.
• While the Purge switch is pressed, the volume delivered display shows the bolus
volume delivered. When releasing the Purge switch, the display shows the sum
of the volume delivered. (The bolus volume delivered is added to the sum of the
volume delivered.)
• When bolus infusion during solution delivery is completed, solution delivery starts
at the same flow rate as before the bolus infusion.
• A Completion alarm occurs if the volume delivered reaches the VTBI when bolus
infusion is used during solution delivery. The solution delivery continues at the flow
rate of KVO (Keep Vein Open).
• The dose setting (flow rate) of Hands On Bolus remains set, and is used when the
next Hands On Bolus is performed, until the power is turned off.
• The upper limit of bolus volume can be set. To change this, contact TERUMO
trained service technicians.