98
Troubleshooting
Door alarm
Occurrence
Cause
Action
• LCD: The Door alarm display
flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Stops.
• The door is open.
• Close the door and securely lock it using the door lock
lever. (See page 31.)
Completion alarm
Occurrence
Cause
Action
• LCD: The Completion alarm
display flashes red.
• Operation indicator: Flashes
green
• Buzzer: Activates.
• Pump: Switches to the keep
vein open (KVO) function and
continues solution delivery.
• The volume delivered has
reached the VTBI.
• A normal operation when the VTBI is set.
• Press the Stop switch to stop the KVO solution delivery.
(See page 42.)
Shutdown Notice alarm
Occurrence
Cause
Action
• LCD: The Shutdown Notice
alarm display flashes red.
• Operation indicator: Flashes
red.
• Buzzer: Activates.
• Pump: Stops.
• Battery has no power left.
• Connect to the AC power supply or TERUFUSION
Standard Rack System. (See page 25.)
* When using a rack system, see the instruction manual
of the rack system.