107
Specifications
Safety functions
Remaining volume
detection setting
function
Selects the Nearly Empty alarm time.
Tube clamp function
When the door is opened, the tube clamp automatically clamps the infusion set.
Flow rate range setting
function A
(Soft limits)
The flow rate setting range can be controlled within the “flow rate setting range”
prescribed for each type of infusion set.
• When you try to set a value outside the range controlled by this function, a
message asking for confirmation is displayed. If you confirm the setting, you
can set the value which is outside the controlled range and the display colour
of the value changes.
Flow rate range setting
function B
(Hard limits)
The flow rate setting range can be controlled within the “flow rate setting range”
prescribed for each type of infusion set.
• Values outside the range controlled by this function cannot be set.
Keypad lock function
Disables switch operations. However, the Start switch, the keypad lock function
disable operation (hold down the Back/Mute switch), and the Power switch still
function. When an alarm is activated, the keypad lock is released.
Bolus reduction
function
Reduces the internal pressure of the tube of the infusion set on the lower side
by reverse motor rotation.
Available
Functions
Infusion set installation
guidance function
Displays the installation status of an infusion set and a screen prompting
installation of the infusion set on the LCD.
Infusion set type
display function
Displays the type and brand of the infusion set on the LCD.
Infusion set type select
function
Enables switching of the infusion set to be used by performing the specified
operation after the power is turned on.
Infusion set drip
volume select function
Changes in accordance with the drip volume of the infusion set.
Drip volume display
function
Displays the drip volume of the infusion set on the LCD.
VTBI setting function
Sets the VTBI of dosage of drug solution.
VTBI time setting
function
Sets the VTBI time of dosage of drug solution.
Purge buzzer sound
function
When the Purge switch is being pressed, the buzzer sounds.
Bolus buzzer sound
function
During a bolus, the buzzer sounds.
Volume delivered clear
function
Clears the volume delivered display to zero.
Standby function
By pressing the Back/Mute switch, the pump can be put in standby mode. In
the standby function, the Start Reminder is delayed.
Standby duration time
select function
Changes the time to activate a Start Reminder in the standby mode.
Buzzer volume select
function
Changes the buzzer volume.
Switch operation
buzzer sound function
Buzzer sounds when a switch operation is performed.
Stop transition buzzer
sound function
Buzzer sounds when the Stop switch is pressed during solution delivery.
LCD brightness select
function
Changes the LCD brightness.
Maintenance timer
function
Time to next maintenance check can be set. That information is displayed on
the LCD.
Date and time setting
function
Sets the date and time.