Instruction manual for MH & MH+M
15
Probing down
Same as for probing up.
Bore measurement
Activating the rotary control handle twice in the same
direction allows bore measurement with detection of the
culmination point.
•
When rotating clockwise, the measurement will start
with probing the upper point of the bore.
•
When rotating counterclockwise, the measurement
will start with probing the lower point of the bore.
When an automatic measurement process is started (not necessarily with the
rotary control handle), it is possible to stop it by a short rotation/activation of
the rotary control handle in the opposite direction of the one of the movement
or by pressing one of the keys of the control panel keyboard.
2.6 Handwheel
On manual models, the main handwheel located above the base is the element that allows you
to move the probe during a measurement.
The handwheel can include a fine adjustment system, which allows to adjust
the displacement to the measurement of small elements. A manual MICRO-
HITE that does not include that system can be upgraded at any time with it.
For further information, please contact your local reseller.
This handwheel and the drive system it is linked to have been specially designed to allow the
user to feel the different pressure on the probe and when the measurement is carried out.
2.7 Lever-type
locking ring
The lever-type ring allows to tighten the drive of the measuring head, which can still be slightly
moved.