4.4 Shield Box Operation with Fixture
The fixture operation mode of standard TC-5916AP is set as 'Handler Mode: Off', 'Probe Mode:
Timer' at the time of release. (You can check it with the query command 'SOLENOID MODE?')
You need to configure the fixture operation mode if you are to use it.
4.4.1 Fixture Operation Modes
The handler and the probe of fixture can operate in 3 modes each, and the operation is automatic
according to the open or close of Shield Box.
The operation of fixture can be controlled with commands via RS-232C. Refer to the list of commands
and each command in detail at
5.5.2.2 Handler & Probe Configuration Commands
Table 4-2
Handler and Probe Operation Modes
Operation Mode
Description
Sensor Mode
The operation of fixture is detected by sensors. This mode can prevent damage to either
the product or DUT caused by malfunction.
Timer Mode
The operation runs by timer without using sensors. The waiting time is already set upon
release and can not be changed.
Off Mode
The operation of fixture is off.
It is NOT recommended operating fixture with the Timer Mode.
Without sensors, the lid may close even when the fixture operation is incomplete, thus causing
damage to the product or DUT.
Summary of Contents for TC-5916AP
Page 1: ...TC 5916AP Pneumatic Shield Box User Manual R20190930...
Page 5: ...Revision History v Revision Record Revision Description Date v1 0 Initial Release 2019 09 23...
Page 9: ...PART 1 Getting Started 1...
Page 13: ...PART 2 Hardware Description 5...
Page 18: ...10 Dimension 2 2 Dimension...
Page 19: ...Dimension 11...
Page 25: ...PART 3 Putting into Operation 17...
Page 55: ...PART 4 Maintenance 47...
Page 63: ...Expendables and Accessories 55 Figure 7 1 TC 5916AP Expendables Location...