40
RS-232C Commands
5.5.3 Lid Operation Commands
OPEN
CLOSE
LID?
OPEN
A command to open the lid.
Type
Command
Input
OPEN
Response
OK
●
OK : The lid opens (or is already open)
If the lid does not open while the 'OPEN' command is sent and 'OK' response comes back,
check the compressed air supply.
CLOSE
A command to close the lid.
The Lid-Close Sensor on a cylinder will glow red if the lid is closed regularly.
Type
Command
Input
CLOSE
Response
{ READY | ERR20 | ERR30 | ERR60 }
●
READY : Lid closed. Ready to test.
●
ERR20 : Only occurs when the Handler is in Sensor Mode. Inward Handler Sensor
detection error. (The horizontal operation of the fixture is incomplete.)
●
ERR30 : Only occurs when the Probe is in Sensor Mode. Downward Probe Sensor
detection error. (The vertical operation of the fixture is incomplete.)
●
ERR60 : Safety error. Multiple operation error.