38
RS-232C Commands
5.5.2.2 Handler & Probe Configuration Commands
SOLENOID MODE?
HANDLER MODE SENSOR
HANDLER MODE TIMER
HANDLER MODE OFF
PROBE MODE SENSOR
PROBE MODE TIMER
PROBE MODE OFF
SOLENOID MODE?
A query command to see the Handler & Probe operation modes.
Type
Query
Input
SOLENOID MODE?
Response
Handler:<mode>, Probe:<mode>
●
<mode>: {OFF | TIMER | SENSOR }
Example
// Below is how to change the Probe Mode from Off to Timer.
SOLENOID MODE?
Handler:OFF, Probe:OFF
PROBE MODE TIMER
OK
It is NOT recommended operating fixture with the Timer Mode.
Without sensors, the lid may close even when the fixture operation is incomplete, thus
causing damage to the product or DUT.
HANDLER MODE SENSOR
A command to set the Handler in Sensor Mode.
Type
Command
Input
HANDLER MODE SENSOR
Response
OK
Summary of Contents for TC-5916AP
Page 1: ...TC 5916AP Pneumatic Shield Box User Manual R20190930...
Page 5: ...Revision History v Revision Record Revision Description Date v1 0 Initial Release 2019 09 23...
Page 9: ...PART 1 Getting Started 1...
Page 13: ...PART 2 Hardware Description 5...
Page 18: ...10 Dimension 2 2 Dimension...
Page 19: ...Dimension 11...
Page 25: ...PART 3 Putting into Operation 17...
Page 55: ...PART 4 Maintenance 47...
Page 63: ...Expendables and Accessories 55 Figure 7 1 TC 5916AP Expendables Location...