P6.0: Temperature unit
P7.0: Real time clock
P8.0: Reset
To enter SETUP mode, hold down “
SET
”key >1 second while in normal
measurement mode. Press “
p
“or “
q
” to select parameters and press
“
ENTER
”for parameter settings. To exit any setting, press “
Esc
” key.
Note: P1.0 memory transmitting function is not available in this model.
P2.0 Memory Clear (CLr)
To clear the stored data, press “
MODE
” key to select the parameter (pH or mV)
which you want to clear before entering setup mode.
While in setup mode, press “
p
“ key to select memory clear function P2.0,
“
CLr
” icon appears on LCD (Fig.6)
Press “
ENTER
” key to enter P2.1.The default “ NO” icon flashes on LCD. Press
“
p
“ key to change and then press “
ENTER
” key to confirm.
NOTE: This procedure will clear 99 memories at a time. Please consider
carefully before delete.
P3.0 Electrode-for pH probe only (ELE)
To view the pH electrode data (slope & offset value) from WT-80:
Press “
MODE
” key to select the mode as pH before entering setting mode.
Press “
p
“ to select P3.0 while in setting mode, “
ELE
“ appears on the LCD.
Press “
ENTER
” key to enter P3.1, LCD displays 1of 4 available slope values
(P3.1 to P3.4). If the value is <75% or >115%, suggest to change electrode
immediately.
Press “
ENTER
” key to review P3.2, P3.3 & P3.4 (Fig. 7)
Press “
ENTER
” to enter P3.5 to view the offset value.(Fig. 8) Offset value is the
mV value of pH 7 and the default offset value is 0.0. The offset value will be
different after re-calibration. When the offset value is out of range +60mV,
strongly suggest you to replace with a new probe.
NOTE: P3.1, 3.2, P3.3 & P3.4 slope definition of buffer:
P3.1
P3.2
P3.3
P3.4
NIST
0.00~4.01
4.01~6.86
6.86~9.18
9.18~14.00
NON-NIST
0.00~4.50
4.50~7.00
7.00~9.50
9.50~14.00