Specification data

All accuracy statements are for one year.

Specification - General

Specification - Temperature ranges

Specification - Resistance ranges

Specification - Electrical connectors (A2) 

Languages

English [Default]

Operating 
temperature

14 … 122°F (-10 … 50°C)

Storage

 

temperature

-4 … 158°F (-20 … 70°C)

Humidity

0 to 90% without condensation 

 

(Def Stan 66-31, 8.6 cat III)

Shock/Vibration

BS EN 61010:2001; Def Stan 66-31, 8.4 cat III

EMC

BS EN 61326-1:1998 + A2:2001

Safety

Electrical - BS EN 61010:2001; CE Marked

Size (L: W: H)

7.1 x 3.3 x 2.0 in (180 x 85 x 50 mm)

Weight

14 oz (400 g)

Power supply

3 x AA alkaline batteries 

Duration
(Measure)

RTD, 

Ω: ≈

 70 hours 

 

mA:

 ≈

 35 hours

mA:

 ≈

 10 hours (24 V Source at 12 mA)

Duration
(Supply)

RTD, 

Ω: ≈

 65 hours 

 

RTD type

Standard

Range °F

Range °C 

Accuracy °F *

Accuracy °C *

Pt50 (385)

IEC 751

-328 ... 1562 

-200 ... 850

0.90

0.50

Pt100 (385)

IEC 751

-328 ... 1562

-200 ... 850

0.45

0.25

Pt200 (385)

IEC 751

-328 ... 1562

-200 ... 850

1.08

0.60

Pt500 (385)

IEC 751

-328 ... 1562

-200 ... 850

0.72

0.40

Pt1000 (385)

IEC 751

-328 ...   752

-200 ... 400

0.36

0.20

D 100 (392)

JIS 1604-1989

-328 ... 1202

-200 ... 650

0.45

0.25

Ni 100

DIN 43760

-76 ...   482

-60 ... 250

0.36

0.20

Ni 120

MINCO 7-120

-112 ...   500

-80 ... 260

0.36

0.20

*Temperature coefficient:
 14 … 50°F, 86 … 122°F

= 0.0028% FS / °F

(-10 … 10°C, 30 … 50°C

= 0.005% FS / °C)

Range (

Ω

)

Excitation (mA)

Accuracy (

Ω

)*

0 ...   400 0.15 ... 0.5 

0.15 

0 ...   400 0.50 ... 3.0 

0.10 

400 ... 1500 0.05 ... 0.8 

0.50 

1500 ... 3200 0.05 ... 0.4 

1.00 

3200 ... 4000 0.05 ... 0.3 

1.30 

*Temperature coefficient:
14 … 50°F, 86 … 122°F = 0.0028% FS / °F
(-10 … 10°C, 30 … 50°C = 0.005% FS / °C)

Range (Measure)

0 to ±55 mA

Accuracy

0.02% of r 3 counts

Temperature coefficient 
 14 … 50°F, 86 … 122°F

 

(-10 … 10°C, 30 … 50°C)

 0.0011% FS / °F

 

(0.002% FS / °C)

Switch detection

Open and closed. 2 mA current.

Loop power output

24 V ± 10%

HART

®

 resistor

250 

Ω 

(menu selection

)

Connectors (A2)

Three 0.16 in (4 mm) sockets

Test Equipment Depot - 800.517.8431

99 Washington Street, Melrose, MA 02176 

TestEquipmentDepot.com  

Summary of Contents for Druck DPI 811

Page 1: ...Druck DPI 811 812 RTD calibrator and RTD loop calibrator User manual K345 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...K345 Issue 4 A1 A2 DPI 812 A3 10 1 3 4 2 5 6 7 9 8 A2 A3 11 B1 12 10 14 15 13 20 21 22 25 17 18 26 19 27 16 23 24 B1 Customer service Visit our web site www gesensing com...

Page 3: ...s uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure Module UPM The DPI...

Page 4: ...ion above it on the display Item 25 Example Settings 8 Display Refer to A3 9 SENSOR PC Communications port Use to connect a Universal Measurement Module UMM or a RS232 cable 10 RTD connectors To measu...

Page 5: ...se this instrument together with a HART communicator to set up and calibrate HART devices Scale To select the applicable international temperature scale IPTS 68 or ITS 90 To select and set up the back...

Page 6: ...select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Additional data S...

Page 7: ...ional Additional data Table 6 Select Settings Step To select and set up values for the Defined Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings Options If...

Page 8: ...adjust the Set Up Table 1 2 Select the RTD or Ohms input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Figure 1 Example configuration To measure the temperature or resis...

Page 9: ...ransmitter calibration DPI 812 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task...

Page 10: ...the top right hand corner 3 Supply the output values to the system Table 7 Example Nudge output a Use Edit to set a value less than the switch value b Use the buttons to change the value in small incr...

Page 11: ...Settings Table 4 5 2 If necessary do a zero correction Table 4 To measure pressure with another operation Figure 8 use the same procedure Figure 8 Example configuration To measure pressure and temper...

Page 12: ...set the new calibration date for the instrument Calibration Procedures Ohms input 1 Let the equipment get to a stable temperature minimum 5 minutes since the last power on 2 Use the calibration menu...

Page 13: ...nt get to a stable temperature minimum 5 minutes since the last power on 3 Use the calibration menu Table 9 to do a three point calibration FS Zero and FS The display shows the applicable instructions...

Page 14: ...850 0 90 0 50 Pt100 385 IEC 751 328 1562 200 850 0 45 0 25 Pt200 385 IEC 751 328 1562 200 850 1 08 0 60 Pt500 385 IEC 751 328 1562 200 850 0 72 0 40 Pt1000 385 IEC 751 328 752 200 400 0 36 0 20 D 100...

Reviews: