HANDLER/SCAN/
INTERFACE
37
Handler Overview
Background
The Handler Interface is used to help bin components
based on the Go-NoGo comparator function test. There
are 6 TTL outputs and one TTL input. The Handler
interface can only be used with the resistance
measurement and compare measurement modes.
Interface and pin
assignment
9 Pin D-SUB
(Female)
Pin Assignment
START
Start the trigger for a single
measurement. Negative edge trigger.
READY
High when the measurement has
finished. The instrument is ready for
the next trigger..
EOT
High when the AD conversion has
completed. The DUT is ready to be
changed.
LOW
High when the compare result is
deemed LO.
HIGH
High when the compare result is
deemed HI.
FAIL
High when the compare result is either
HI or LO (fail).
PASS
High when the compare result is GO
(pass).
Note
The output current from all the terminals and the +5V
terminal cannot exceed 60mA.