Preparing for first-time use
50 Testboy
®
TV 470
Setting up the test standard
In order to make changes to the test standard settings, press the "NORM"
key in the main menu.
The display changes according to the pre-sets. Refer to VDE701/702 to EN
62353 (VDE0751-1).
By choosing standard EN 62353 (VDE0751-1) (Medical electrical equip-
ment) an additional menu with the following setting options is displayed:
Selection menu EN 62353 (VDE751-1)
Protection class
Protection class-I or II can be set.
Type of measurement
Selection of how leakage currents are measured, either by "direct measure-
ment" or by "equivalent measurement".
Application type
Here the necessary application type B, BF or CF can be set.