Preparing for first-time use
Testboy
®
TV 470
53
ENGLISH
Connecting and using the bar code scanner.
Connect the bar code scanner to the rear of the TV 470. An acoustic signal
indicates that the scanner is ready.
In order to read a bar code, aim the scanner at the bar code and press the
scan key. An acoustic signal indicates that the code has been read success-
fully.
Repeat tests are simplified when bar codes are used.
In "Auto" mode, reading a bar code will automatically start a measurement.
All measurement values will be saved along with the bar code. With the
analysis, all measurements as well as the bar code will be saved in the da-
tabase.
With a repeat test, all results are automatically sorted to the DUT. For this
reason, ensure that each bar code is assigned only once.
To start a measurement
In order to start a measurement, press the key with the function "Measure"
or read a bar code in the "Auto" mode. In "Auto" mode, the device starts with
the first measurement.
In manual mode, the measurement choices are displayed.
The "Auto" measurement sequence follows the order given by the directive.
This is dependent on the protection class setting on the device.
Protection class-I:
PE conductor resistance – insulation resistance – leakage current – equiva-
lent leakage current
Protection class-II:
Insulation resistance – leakage current – equivalent leakage current
Protection class-III:
Insulation resistance
When testing devices of protection class-I and II, the measurement
can be stopped after the leakage current measurement by pressing
Summary of Contents for TV 470
Page 1: ...TV 470 Version 1 1...
Page 17: ...Inbetriebnahme Testboy TV 470 17 DEUTSCH Bild 1 Bild 2...
Page 20: ...Inbetriebnahme 20 Testboy TV 470 Bild 5 Bild 6...
Page 55: ...Preparing for first time use Testboy TV 470 55 ENGLISH Illustration 1 Illustration 2...
Page 58: ...Preparing for first time use 58 Testboy TV 470 Illustration 5 Illustration 6...
Page 93: ...Ingebruikneming Testboy TV 470 93 NEDERLANDS Afb 1 Afb 2...
Page 96: ...Ingebruikneming 96 Testboy TV 470 Afb 5 Afb 6...
Page 131: ...Prepara o para a primeira utiliza o Testboy TV 470 131 PORTUGU S Figura 1 Figura 2...
Page 134: ...Prepara o para a primeira utiliza o 134 Testboy TV 470 Figura 5 Figura 6...
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......