Information about DIN EN 62353 (VDE 0751-1:2008-8)
68 Testboy
®
TV 470
Information about DIN EN 62353 (VDE 0751-1:2008-8)
(Retest and inspection after overhauling of medical electrical devices)
The relevant original valid directive is decisive for the implementa-
tion of tests or determining limit values!
The tests should be carried out in the order specified:
|
Visual inspection
|
PE conductor resistance
|
Insulation resistance (when required by the manufacturer)
|
Leakage current
|
Functional test and documentation
PE conductor resistance
The resistance and integrity of the PE conductor should be measured.
Measurement voltage 4 – 24 V,
Measurement current > 200 mA
Limits
< 0.3
Ω
inclusive of the mains lead
With removable leads where applicable 0.2 or 0.1
Ω
Important notes
|
Move the connection cables during measurements
|
As the test lead resistance is part of the measurement, the test lead
should have a good electrical connection.
Summary of Contents for TV 470
Page 1: ...TV 470 Version 1 1...
Page 17: ...Inbetriebnahme Testboy TV 470 17 DEUTSCH Bild 1 Bild 2...
Page 20: ...Inbetriebnahme 20 Testboy TV 470 Bild 5 Bild 6...
Page 55: ...Preparing for first time use Testboy TV 470 55 ENGLISH Illustration 1 Illustration 2...
Page 58: ...Preparing for first time use 58 Testboy TV 470 Illustration 5 Illustration 6...
Page 93: ...Ingebruikneming Testboy TV 470 93 NEDERLANDS Afb 1 Afb 2...
Page 96: ...Ingebruikneming 96 Testboy TV 470 Afb 5 Afb 6...
Page 131: ...Prepara o para a primeira utiliza o Testboy TV 470 131 PORTUGU S Figura 1 Figura 2...
Page 134: ...Prepara o para a primeira utiliza o 134 Testboy TV 470 Figura 5 Figura 6...
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......