Information about DIN EN 62353 (VDE 0751-1:2008-8)
Testboy
®
TV 470
71
ENGLISH
Equivalent patient leakage current
Allowable values from long-term flow equiva-
lent patient leakage current
Value in mA
Application part
B
BF
CF
Equivalent patient leakage current.
–
5.0
0.05
Note 3
: Mobile X-ray machines with mineral insulation.
Important notes
|
Equivalent device leakage current insulation test is only valid when all
circuits in the device are activated.
|
The application part type determines the limit value.
If all safety-relevant parts are not reached with the equivalent device leakage
current, then a leakage current measurement should be made instead. Use
either the direct method or the indirect method
. With the indirect method,
the DUT has to be in an isolated position.
Summary of Contents for TV 470
Page 1: ...TV 470 Version 1 1...
Page 17: ...Inbetriebnahme Testboy TV 470 17 DEUTSCH Bild 1 Bild 2...
Page 20: ...Inbetriebnahme 20 Testboy TV 470 Bild 5 Bild 6...
Page 55: ...Preparing for first time use Testboy TV 470 55 ENGLISH Illustration 1 Illustration 2...
Page 58: ...Preparing for first time use 58 Testboy TV 470 Illustration 5 Illustration 6...
Page 93: ...Ingebruikneming Testboy TV 470 93 NEDERLANDS Afb 1 Afb 2...
Page 96: ...Ingebruikneming 96 Testboy TV 470 Afb 5 Afb 6...
Page 131: ...Prepara o para a primeira utiliza o Testboy TV 470 131 PORTUGU S Figura 1 Figura 2...
Page 134: ...Prepara o para a primeira utiliza o 134 Testboy TV 470 Figura 5 Figura 6...
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......