Information about DIN EN 62353 (VDE 0751-1:2008-8)
72 Testboy
®
TV 470
Device leakage current
With devices where you cannot determine if all mains affected components
are tested with the replacement device leakage current measurement or the
replacement device leakage current measurement for another reason can
not be carried out, then the device leakage current or the differential current
measurement can be measured.
Allowable values from long-term flow device
leakage current
Value in mA
Application part
B
BF
CF
Device leakage current generally
0.5
0.5
0.5
Leakage current for devices according to notes 1
and 3
2.5 2.5 2.5
Leakage current for devices according to note 2
5.0
5.0
5.0
Device leakage current for protection class-II de-
vices and protection class-I devices, which are not
connected to the PE conductor conductive com-
ponents.
0.1 0.1 0.1
Patient leakage current
Allowable values from long-term flow equiva-
lent leakage current
Value in mA
Application part
B BF CF
Direct current
0.01
–
–
Patient leakage current
Alternating current
0.10
–
–
Summary of Contents for TV 470
Page 1: ...TV 470 Version 1 1...
Page 17: ...Inbetriebnahme Testboy TV 470 17 DEUTSCH Bild 1 Bild 2...
Page 20: ...Inbetriebnahme 20 Testboy TV 470 Bild 5 Bild 6...
Page 55: ...Preparing for first time use Testboy TV 470 55 ENGLISH Illustration 1 Illustration 2...
Page 58: ...Preparing for first time use 58 Testboy TV 470 Illustration 5 Illustration 6...
Page 93: ...Ingebruikneming Testboy TV 470 93 NEDERLANDS Afb 1 Afb 2...
Page 96: ...Ingebruikneming 96 Testboy TV 470 Afb 5 Afb 6...
Page 131: ...Prepara o para a primeira utiliza o Testboy TV 470 131 PORTUGU S Figura 1 Figura 2...
Page 134: ...Prepara o para a primeira utiliza o 134 Testboy TV 470 Figura 5 Figura 6...
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......