Appendix
Testboy
®
TV 470
75
ENGLISH
DIN EN 62353 (VDE 0751-1:2008-8)
|
Allowable leakage current values
|
Allowable values for long-term substitute-, device-, and patient leakage
current.
Application part
Type B
Type
BF
Type
CF
Device leakage current generally
0.5 mA
0.5 mA
0.5 mA
Device leakage current for devices ac-
cording to notes 1 and 3
2.5 mA
2.5 mA
2.5 mA
Leakage current for devices according
to note 2
5.0 mA
5.0 mA
5.0 mA
Device leakage current for protection
class-II devices as well as exposed
conductive components from class-I
devices which are not connected to the
PE conductor.
0.1 mA
0.1 mA
0.1 mA
Equivalent device leakage current for
protection class-II devices as well as
exposed conductive components from
class-I devices which are not con-
nected to the PE conductor.
|
In the PE conductor or
parts connected to the
PE conductor
|
Devices with mineral in-
sulation and devices ac-
cording to note 1
|
According to note 2
0.2 mA
1.0 mA
5.0 mA
10.0 mA
0.2 mA
1.0 mA
5.0 mA
10.0 mA
0.2 mA
1.0 mA
5.0 mA
10.0 mA
Mobile X-ray machines with additional
PE conductor
5.0 mA
5.0 mA
5.0 mA
Mobile X-ray machines without addi-
tional PE conductor
2.0 mA
2.0 mA
2.0 mA
Summary of Contents for TV 470
Page 1: ...TV 470 Version 1 1...
Page 17: ...Inbetriebnahme Testboy TV 470 17 DEUTSCH Bild 1 Bild 2...
Page 20: ...Inbetriebnahme 20 Testboy TV 470 Bild 5 Bild 6...
Page 55: ...Preparing for first time use Testboy TV 470 55 ENGLISH Illustration 1 Illustration 2...
Page 58: ...Preparing for first time use 58 Testboy TV 470 Illustration 5 Illustration 6...
Page 93: ...Ingebruikneming Testboy TV 470 93 NEDERLANDS Afb 1 Afb 2...
Page 96: ...Ingebruikneming 96 Testboy TV 470 Afb 5 Afb 6...
Page 131: ...Prepara o para a primeira utiliza o Testboy TV 470 131 PORTUGU S Figura 1 Figura 2...
Page 134: ...Prepara o para a primeira utiliza o 134 Testboy TV 470 Figura 5 Figura 6...
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......