Chapter 4 – Using the F4T Front Panel
Page 30
TestEquity 101H Temperature/Humidity Chamber
Jump:
Repeats previous steps in the profile starting at the step set with Jump to Step for the
Number of Times set. This option is not available for step 1.
•
Jump Step
•
Jump Count
•
Power, Humidity, Purge, Event 4 through 8
(On/Of functions)
End:
Sets the Temperature or Humidity condition after the profile is complete.
•
User
(program will end with the same Temperature and Humidity Static Set Point values and
Event settings that were entered before the profile was run). If you want to select
User
, you
should make sure that the Temperature and Humidity Static Set Point values and Event
settings are indeed what you want the chamber to be at when the profile ends.
•
Off
(the Temperature and Humidity control mode is set to Off –
Not Recommended, see
Note Below!
)
•
Hold
(the Temperature and Humidity Set Point is set to the last value used in the profile).
This may be a preferable choice for most applications, since you can define the desired
chamber conditions in the next-to-last step in the profile.
NOTE about selecting an End step of Off:
If you select an End step of
Off
, this sets the
Control Mode
in the
Loop Menu
to
Off
. This
results in the inability to set a
Static (Manual) Set Point
. You would need to enter the
Loop
Menu
and change
Control Mode
to
Auto
. See “Changing Loop Operational Parameters” for
information on how to access this menu.