Chapter 4 – Using the F4T Front Panel
Page 32
TestEquity 101H Temperature/Humidity Chamber
•
Press
Add Step
. You will now see
Step 4
.
o
Change step type to
Ramp Time
.
o
Enter
-40
for
Target Set Point 1
.
o
Enter
2
for
Hour
and
0
for
Minutes
.
o
Press
Done
.
•
Press
Add Step
. You will now see
Step 5
.
o
Change step type to
Soak
.
o
Enter
2
for
Hour
and
30
for
Minutes
.
o
Press
Done
.
•
Press
Add Step
. You will now see
Step 6
.
o
Change step type to
Ramp Time
.
o
Enter
85
for
Target Set Point 1
.
o
Enter
2
for
Hour
and
0
for
Minutes
.
o
Press
Done
.
•
Press
Add Step
. You will now see
Step 7
.
o
Change step type to
Jump
.
o
Enter
3
for
Jump Step
. Enter
9
for
Jump Count
.
o
Press
Done
.
•
Press
Add Step
. You will now see
Step 8
.
o
Change step type to
Instant Change
.
o
Enter
23
for
Target Set Point 1
.
o
Enter
0
for
Hour
and
0
for
Minutes
.
o
Select
Guaranteed Soak Enable 1
to
On
.
o
Press
Done
.
•
Press
Add Step
. You will now see
Step 9
.
o
Change step type to
End
.
o
Press
Profile End Step Loop Action 1
, press
User
and change it to
Hold
.
o
At
Power
press
On
and change it to
Off
.
o
Press
Done
.
•
Press
Options
.
o
Press
Done
to save the profile.
or…
o
Press
Run Profile
to run the profile.