background image

 

Chapter 8 – Maintenance 

 

 

Page 8-4 

TestEquity 123C Temperature Chamber 

How to Verify the Performance

 

 

These tests verify the performance of the heating, cascade refrigeration mode, electrical controls, 

Temperature Controller, and air circulation systems. The chamber should meet all published 

performance specifications if all of these tests are successfully passed. These tests assume that 

the Temperature Controller’s setup and tuning values have not been changed from the values as 

shipped from TestEquity. Also, the EZ-Zone Limit Controller high limit must be set to over 

+85°C (+88°C would be fine), and the low limit set to –75°C. If the chamber fails any of these 

tests, it should be removed from service to prevent further damage until the cause of the problem 

is determined and resolved. 

 

1.

 

The chamber interior should be empty and at ambient temperature, approxi23°C. 

2.

 

Turn ONLY the Power Switch ON. 

3.

 

Set the F4 Controller Set Point to +85°C. Then, turn the Conditioning Switch ON. 

4.

 

The Heat Light should be ON continuously and the Cool light should be OFF. 

5.

 

The chamber should heat up to about +80°C and begin controlling (HEAT Light cycles 

ON/OFF) within approximately 12 minutes. 

6.

 

The chamber temperature should slowly increase and stabilize to +85°C. It should NOT 

overshoot 85°C by more than a few tenths of a degree, and the compressors should 

NOT need to turn ON in order to ma85°C. 

7.

 

After stabilization, the chamber temperature should vary no more than ±0.2°C, or a total of 

0.4°C. 

8.

 

Let the chamber stay at +85°C for two hours. 

9.

 

After two hours at +85°C, set the Temperature Controller Set Point to –65°C. 

10.

 

The high-stage compressor should turn ON within a few seconds. Then, the low-stage 

compressor should turn ON within approximately 30 seconds, the Cool Light should be ON 

continuously and the Heat Light should be OFF. 

11.

 

The chamber should cool down to about –60°C and begin controlling (Cool Light cycles 

ON/OFF) within 1 hour and 40 minutes (100 minutes). 

12.

 

The chamber temperature should slowly decrease and stabilize to –65°C. It should NOT 

undershoot beyond –65°C by more than a few tenths of a degree, and the compressors should 

NOT need to turn OFF in order to maintain –65°C. 

13.

 

After stabilization, the chamber temperature should vary no more than ±0.5°C, or a total of 

1°C.  

14.

 

Set the Temperature Controller Set Point to +23°C. The chamber should begin to heat up. 

The compressors should turn off within approximately 1 minute. 

15.

 

This concludes the chamber performance verification tests. 

16.

 

Let the chamber heat up to +23°C before turning the Conditioning Switch OFF. 

 

Summary of Contents for 123C

Page 1: ...1 2018 Model 123C Temperature Chamber Serial Number 330218 and above Operation and Service Manual TestEquity LLC 6100 Condor Drive Moorpark CA 93021 Support 877 512 3457 Toll Free 805 480 0638 Corpora...

Page 2: ......

Page 3: ..._________________________________ 3 4 Internal Test Fixtures ___________________________________________________________________ 3 4 Chapter 4 Limit Controller_________________________________________...

Page 4: ...____________8 7 Troubleshooting _________________________________________________________________8 9 Refrigeration System Charging Instructions_________________________________________8 11 R 404A High...

Page 5: ...ard symbol 2 precedes an electric shock hazard CAUTION or WARNING statement Installation Safety Notices 2 WARNING The power cord is equipped with a NEMA 5 20P grounded polarized plug To prevent a shoc...

Page 6: ...safe limits This could occur if your test sample exceeds the live load rating of the chamber or if the chamber s refrigeration system fails This chamber has a set of contacts that can be used to remov...

Page 7: ...e the retainer Keep in mind that there is also Velcro that holds the retainer in place Place the retainer in the front of the pallet on the floor 5 Locate the two ramps under the chamber and remove th...

Page 8: ...nment Operation above 30 C 85 F or below 16 C 60 F ambient room temperature is NOT recommended Condensate Drain The chamber has a condensate drain connection on the rear of the chamber This provides a...

Page 9: ...ontroller automatically turns the refrigeration system on or off as required based on the deviation from temperature set point Refer to the separate F4T or F4 Temperature Controller manual for details...

Page 10: ...t 1 Mode When the CONDITIONING Switch is in the EVENT 1 position you can enable and disable all chamber functions through the Power button on the F4T Temperature Controller or Event 1 Digital Output 1...

Page 11: ...cts and surfaces that are hot or cold to prevent severe burns or frostbite Protective gloves are recommended CAUTION If your test sample is energized it may be capable of raising the workspace tempera...

Page 12: ...ing low temperature operation or water when maintaining over 0 C and cooling is required To avoid moisture condensation make sure the port plugs are inserted at all times Also avoid opening the chambe...

Page 13: ...d is written by the manufacturer Watlow for a wide variety of applications and configurations Not all features or functions are applicable Only the capabilities of a model PM6L1AJ AAAABAA are applicab...

Page 14: ...your actual chamber temperature set point to prevent nuisance tripping 5 Press the RESET Key to return to the Home Page 6 The upper display will show the actual chamber temperature while the lower dis...

Page 15: ...ower to your test sample if the Limit Controller s temperature limits are exceeded The safety contacts are rated as follows Resistive 10 A 250 VAC or 10 A 28 VDC Inductive 7 A 250 VAC To access the sa...

Page 16: ......

Page 17: ...2 flowing GN2 Gaseous Nitrogen Installation Option TE 1239 Connect a supply of GN2 with a maximum pressure of 100 psig to the 1 4 inch FPT fitting which is designated PURGE on the rear panel Dry Air I...

Page 18: ...adjust the flow of purge gas into the chamber The flowmeter has a scale calibrated in SCFM The flow of purge gas should be adjusted to the minimum amount required to obtain the desired drying in the c...

Page 19: ...rating properly by following the procedure in How to verify the chamber performance How can I modify the chamber to cool faster Unfortunately there is little you can do to improve upon the designed in...

Page 20: ...s the ability of the temperature controller to accurately display a temperature measurement when compared to a standard The controller display accuracy is 1 66 C However the total measurement accuracy...

Page 21: ...u would like to use is qualified Can Should I put a filter in front of the condenser air inlet No TestEquity does not recommend this Just follow the maintenance procedures and clean the condenser fins...

Page 22: ......

Page 23: ...min 53 min ultimate 65 C 15 min 21 min 32 min 45 min 54 min ultimate Note Transition times are measured after a 2 hour soak at the start temperature with an empty chamber Does not include the effect...

Page 24: ......

Page 25: ...only Preventive Maintenance Intervals Daily or As Needed Clean chamber interior and exterior Listen for abnormal noise or vibration Every 3 Months Inspect the door seal Inspect the refrigeration machi...

Page 26: ...eing able to recognize changes from normal operating noises can be a valuable way to identify problems and prevent further damage Examples of noises to be aware of include Circulator motor and fan noi...

Page 27: ...frigerant line which could break 1 Unplug the chamber from the power source 2 Remove the side and rear panels 3 Inspect for signs of water leaks 4 Inspect for signs of refrigeration tubing abrasion 5...

Page 28: ...er temperature should slowly increase and stabilize to 85 C It should NOT overshoot beyond 85 C by more than a few tenths of a degree and the compressors should NOT need to turn ON in order to maintai...

Page 29: ...ar terminals and burned or excessively pitted contacts on contactors How to Clean the Condenser 1 Disconnect the chamber from the power source 2 Remove the lower front panel 3 Clean the condenser and...

Page 30: ...hamber and adjusting the temperature controller s calibration to correct for these errors is incorrect The correct way to adjust what the temperature controller displays compared to what is measured a...

Page 31: ...ght glass 7 to the thermostatic expansion valve 20 The thermostatic expansion valve controls the feed of liquid refrigerant to the evaporator circuit of the cascade condenser and reduces the pressure...

Page 32: ...balance pressure when the system is off within safe limits Both the low and high stages each have a high pressure switch 8A 8B respectively which turns off the entire refrigeration system in the event...

Page 33: ...both compressors are OFF 2 Setpoint is low than chamber temperature but F4T Compressor Output F4 DigitalOut 8 is not ON 3 If the Cool light is ON compressors are ON 1 Pressure switch PS1 or PS2 is tr...

Page 34: ...uity and only occurs in cool mode 1 Re enter values as documented in the controller manual NEVER use the Autotune feature 2 Control parameters may need to be changed for your unique test conditions 3...

Page 35: ...performance as outlined in How to verify the chamber performance NOTE If the F4 Controller is cycling Cool Light cycles on off the sightglass may appear 1 2 to 2 3 full This is normal R 508B Low Stage...

Page 36: ...out of service for even one day then you should purchase critical spare parts in advance Although most parts are standard and available from a variety of local distributors some parts are either harde...

Page 37: ...0T US DC24 SSR1 200177 1 ea Relay Solid State 10A 5VDC in Omron G3NE 210T US DC5 SSR4 6 200129 2 ea Switch Carlingswitch LRA211 RA B 125N SW1 3 200023 2 ea Switch ON OFF ON Carlingswitch RC911 RA B 0...

Page 38: ...on Cooling Danfoss 068U202700 11 100314 1 ea Filter Drier Danfoss 023Z5048 6 34 100524 2 ea High Pressure Cutout Johnson Controls P20DB 1D 8A 8B 100341 2 ea Hot Gas Bypass Regulator Sporlan ADRI 1 4 0...

Page 39: ...Limit Sides both Alt high High only or LoW Low only L hY Limit Hysteresis 2 Change not recommended A ty Alarm Type oFF Do Not Change PAr1 Upper Display ACPu Alt none PAr2 Lower Display Lst Alt Lh s Hi...

Page 40: ......

Page 41: ...bor if they initiate service on their own without prior approval from TestEquity Replacement parts must be supplied by TestEquity for in warranty claims Purchaser will not be reimbursed for parts they...

Page 42: ......

Page 43: ...Chapter 10 Drawings TestEquity 123C Temperature Chamber Page 10 1 Chapter 10 Drawings...

Page 44: ......

Page 45: ...Electrical Subpanel Component Location...

Reviews:

Related manuals for 123C