Operation manual Pneumator
∙
Testo Industrial Services
∙
Version 1.4
∙
2019
∙
Page 9
Operation / application:
After selecting the operation mode CTRL and leaving the menu (MENU button) a
zero adjustment is automatically conducted, in case AUTO ZERO = ON.
Subsequently, the reference sensor is venting („VENT “).
The respective pressure set point value now results now from the nominal pressure
value (see upper display middle, initial value = full measuring range), multiplied by a
percentage value (see lower display middle, initial value = 0%). The respective set
point value is indicated in the lower right display, the unit in the upper right. Note: The
new set point value is activated immediately after the change of the absolute value or
of the percental value.
Example
(compare illustration):
- Nominal value: 100.00 Pa
- Percental value: 0%
- Current set point value = 100 Pa * 0% = 0 Pa
You can switch between the absolute value and the
percental value by pressing OK.
Change of the percental value (lower display middle - framed bold):
For the gradual up - and down regulation of the pressure value, e.g. 0 / 25 / 50 / 75 /
100 % of the absolute value. The step size x is defined in the submenu „range“ .
UP:
Pressure rise of x % of the measuring range
DOWN:
Pressure reduction of x % of the measuring range
0%:
Pressure = 0 Pa
100%:
Pressure = 100% of the measuring range
Change of the absolute value (upper display center - framed bold):
For the start-up of a given pressure value, e.g. 610 hPa. In most cases it is advisable
to leave this value on 100% of the measuring range.
UP:
Digit +1
DOWN: Digit
-1
0%:
one digit position to the left
100%:
one digit position to the right
Retrieval of the calibration data:
As soon as the central display figure indicates a stable value, the respective
calibration data can be retrieved:
- physical unit: Display value upper right
- pressure set value: Display value in the lower right
- actual value of test device: Display of the test device
- Reference value: Display middle
Manual zero balance (ZERO)
In case the reference measuring cell is to be set to zero during the calibration, please
use the ZERO button. After approximately 3-5 seconds, the previous status is started
up again.