2 Specifications
4
Acceleration readings are monitored throughout the duration of the
measurement program and saved when the set limit value is
exceeded.
Data logger programming and measurement report output are
implemented via PDF files, no software needs to be installed.
Product versions T1 and T2 are single-use data loggers with a
time-limited service life.
2.2.
Technical data
testo 184 T1, T2, T3, T4
Feature
Values
Display T1,
T4:
no
T2, T3: yes
Probe type T1, T2, T3: internal NTC temperature sensor
T4: internal PT1000 temperature sensor
Measuremen
t channels
1 internal
Measure-
ment
parameters
[unit]
Temperature [°C, °F]
Measuring
range
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Accuracy
T1, T2, T3: ±0.5 K
T4: ±0.8 K (-80 to -35.1 °C), ±0.5 K (-35.0 to 70 °C)
Resolution 0.1
°C
Operating
temperature
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Storage
temperature
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Battery type T1: internal, non-replaceable
T2: internal, non-replaceable
T3: CR2450, replaceable
T4: TLH-2450, replaceable