6 Using the product
54
Show flue gas matrix
This function is only available if the measurement parameter
CO
has been activated in the readings display.
Calling up the function:
✓
The flue gas function is opened.
>
[Options]
→
Flue Gas Matrix
:
Options
>
[Options]
→
Clipboard
: Data are saved to the clipboard:
>
[Options]
→
Delete clipboard
: Any data saved to the clipboard
is deleted.
>
[Options]
→
Save
: The readings are saved in a record.
>
[Options]
→
Show Graphic
: The readings are displayed in
form of a line graph.
>
[Options]
→
Show Numerical Values
: Data are displayed as
numerical values.
>
[Options]
→
System Type
: (This function is not available
during a measurement). Set the system type to be able to
configure the ideal zone (green) of the flue gas matrix, using the
limits pre-configured for each system type.
>
[Options]
→
Reset Graphic
: The displayed graphical values
are deleted.
>
[Options]
→
Limits
: (This function is not available during a
measurement). Enter limits to be able to configure the ideal
zone (green) of the flue gas matrix.
>
[Options]
→
CO + O2
or
CO + CO2
: Choose which parameter
should be assigned to the x-axis of the display matrix (O2 or
CO2).
>
[Options]
→
Measurement view
: (This function is not available
during a measurement). Open the readings display menu.
3.
6.2.3.
Draught-Measurement
Calling up the function:
✓
A flue gas probe must be connected.
1.
[ ]
→
Measurements
→
[OK]
→
Draught
→
[OK]
.
Performing the measurement:
The pressure socket of the instrument must be free (i.e.
unpressurized, not closed).
Do not measure for longer than 5 min, as the drift of the
pressure sensor means that the readings could be outside
the tolerance limits.