6 Using the product
62
Performing the measurement:
1.
Start measurement:
[
]
.
- The measurement starts and the measuring value is displayed
graphically (trend display).
- An audible alarm signal is triggered when the alarm limit is
reached.
2.
Quit measurement:
[
]
.
3. Confirm the message:
[OK]
.
Options:
>
[Options]
→
Clipboard
: Data are saved to the clipboard.
>
[Options]
→
Delete clipboard
: Any data saved to the clipboard
is deleted.
>
[Options]
→
Save
: The readings are saved in a record.
>
[Options]
→
A-Thresh.
: The alarm limits menu is opened.
6.2.14.
CO2 ambient
✓
An ambient CO2 probe (0632 1240) must be connected.
In order to obtain correct readings, it is imperative to enter
the prevailing absolute pressure. This can be entered
directly (
Pressure Absolute
), or it is automatically
calculated when entering
Altitude
and barometric pressure
(
Pressure Barom.
).
Calling up the function:
>
[ ]
→
Measurements
→
[OK]
→
CO2 Ambient
→
[OK]
.
Performing the measurement:
1.
Select parameter
→
[Edit]
.
2.
Enter values.
[
▲
]
,
[
▼
]
and partly
[
◄
]
,
[
►
]
→
[OK]
.
3.
Start measurement:
[
]
.
4.
Quit measurement:
[
]
.
- The ambient CO2 value is displayed.
Options:
>
[Options]
→
Clipboard
: Data are saved to the clipboard.
>
[Options]
→
Delete clipboard
: Any data saved to the clipboard
is deleted.
>
[Options]
→
Save
: The readings are saved in a record.