4 Transmitter
10
Response time
t
63
≤
3 s when changing from -75 °C
td
to -30 °C
td
t
90
≤
9 s when changing from -75 °C
td
to -30 °C
td
t
63
≤
300 s when changing from -30 °C
td
to -75 °C
td
t
90
≤
1080 s when changing from -30 °C
td
to -75 °C
td
Resolution
Measuring range
Resolution
0 to 100 % RH
0.001
0.001 to 28 g/kg
0.001
0.01 to 194 gr/lb
0.01
0 to 31 g/m
3
0.001
0.001 to 14 gr/ft
3
0.001
0.1 to 9999.9 ppm
(vol)
0.1
-90 to -20 °C
td
0.1
-130 to -4 °F
td
0.1
-110 to 30 °C
tdA
0.1
-165 to 86 °F
tdA
0.1
-40 to 70 °C
0.01
-40 to 158 °F
0.01
Meas. cycle
• 1/sec
Interface
•
Mini-DIN for P2A software (adjustment and parameterization
software)
Voltage supply
•
4-wire (separate signal and supply lines): 20 to 30 V AC/DC,
max. 150 mA power consumption