4 Transmitter
14
Pos: 26 /TD/Überschriften/MUF/5.2/6.2 Produktbeschreibung @ 3\mod_1234258723551_79.doc @ 24008 @ 2 @ 1
4.2.
Product description
Pos: 27 /TD/Produktbeschreibung/Übersicht/MUF 63xx/Systemkomponenten 6781 @ 3\mod_1239868226476_79.doc @ 31305 @ 3 @ 1
4.2.1.
System components
Application
System components
A With clean air and a process pressure
corresponding to the presetting of the
measurement chamber (1), a measurement
chamber (1) is required for the rapid
assembly (without de-pressurizing the
installation).
•
1: Measurement chamber
• testo
6781
with
G1/2 thread
B With clean air and a process pressure
greater than or less than 7000 hPa (1000 to
35,000 hPa) that therefore does not
correspond to the presetting of the inflow
valve, a measurement chamber (1) with a
flow meter (2) is required.
•
1: Measurement chamber
•
2: Flow meter
• testo
6781
with
G1/2 thread