4 Transmitter
32
4.4.5.2.
Editing Main Menu Alarm
The alarm statuses are shown on the display. You can choose
whether the alarm is to be used to monitor limit values or as a
collective alarm. If an alarm is to be used to monitor limit values,
you can choose between monitoring the minimum or maximum
value and set a limit value and hysteresis for each alarm.
In addition, every alarm can be linked to a clearly visible visual
alarm (display background lighting flashes).
An alarm delay between 0 and 240 seconds can still be assigned to
every alarm used for limit value monitoring so that the visual alarm
is delayed. If the alarm status goes out within the set alarm delay
time, the visual alarm is not triggered. .
With an alarm status present, the visual alarm can be reset by
means of acknowledgement. The triggering of a new alarm cannot
be enabled until after the alarm status goes out.
1.
In the Measuring Mode press
SET
, select
Main Menu Alarm
with
or
and confirm selection with
SET
.
-
Four alarms can be parameterized.
2.
Select
Alarm x
with
or
and confirm selection with
SET
.
Using alarm to monitor limit values
Monitoring minimum
Monitoring maximum
Hysteresis
Hysteresis
Limit value
Limit value
3.
Select
Channel x
(e.g. "Channel 1") with
or
and confirm
selection with
SET
.
4.
Select
Max control
or
Min control
with
or
(see graphic).
5.
Press
SET
and edit
Limit value
as well as
Hysteresis
: Scroll
one digit to the right using
and increase value of digit by 1
using
. Confirm with
SET
or abort entry with
ESC
.
6.
Select
Visual alarm
with
or
. Select
YES
or
NO
with
or
. Confirm with
SET
or abort entry with
ESC
.
7.
Press
SET
and edit
Alarm delay
: Scroll one digit to the right
using
and increase value of digit by 1 using
. Confirm with
SET
or abort entry with
ESC
.
8.
Return to
Channel x
with
ESC
.