4 Transmitter
35
1
Channel 1
test contacts
2
Multimeter
3.
Return to
Test Analog Output
with
ESC
and continue to
Max./min. values
with
or
.
Reading off min./max. values of channels
To reset the max./min. values, see Editing Reset main menu page
39
1.
Read off the min./max. values of the channel in succession with
or
and return to the
Main Menu Analysis
with
ESC
.
2.
Continue to
Main Menu Message
with
or
or return to
Measuring Mode with
ESC
.
Pos: 52 /TD/Produkt verwenden/MUF 63xx/Hauptmenü Meldungen bearbeiten 6781 @ 3\mod_1240905240318_79.doc @ 32203 @ 4 @ 1
4.4.5.5.
Editing Message main menu
Messages can be confirmed/acknowledged, the last messages can
be called up and the display of the messages can be switched on
or off.
1
Message text
2
Additional information on
the message
3
Message number
(example: "4/7" refers to
the fourth of seven
messages)
4
Number of messages
present (example: "4/7"
refers to the fourth of
seven messages)
5
Message code (see
Status, warning and error
messages page 39).
Using the P2A software (see Parameterizing, adjusting and
analyzing software (P2A software) page 47) you can
predefine whether messages are to be shown in the
display.
1.
In the Measuring Mode press
SET
, select
Main Menu Message
with
or
and confirm selection with
SET
.