4 Transmitter
40
In the transmitter, the last 60 status messages and the last 120
error and warning messages are stored in a ring memory. There is
no limit in the P2A software.
Pos: 58 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Statusmeldungen 678x @ 3\mod_1238508288143_79.doc @ 30547 @ 3 @ 1
4.5.1.
Status messages
Status messages show the current operating status of the testo
6781.
Message Display
Description
02506
Sensor
initialization
Message appears while the
transmitter is starting up. If the
message disappears, the
transmitter is ready for operation.
01D19 Service
plug The Mini-DIN socket is connected
to: the USB adapter for P2A
software, the adjustment adapter
or the service plug (is not
recorded/no number)
00300
New limit value
The limit value has been changed
or shifted
00301
Scaling changed
The scaling has been changed
02104
Analog out adjust An analog adjustment has been
made
00500 Transmitter
reset
The transmitter has been
restarted.
0052F Reset
Min/Max
Resets all saved Min/Max values
for all channels
00503
Reset device to
fact
A factory reset of the transmitter
was performed
00307
User Setting
Change
User Setting Change: General
settings were changed for the
transmitter