4.5.3.
Transmitter error messages
Error messages show a current malfunction.
Message Display
Cause
Remedying of fault
01505 Watchdog
error
Due to a processor
error, the transmitter
performs an automatic
restart.
If the problem occurs
frequently, contact
Testo Service.
01115
Low adjustment
temperature
The ambient
temperature is too low
during the pressure
adjustment.
Take necessary
measures to raise
ambient temperature,
e.g. through heating.
01116
High adjustment
temperature
The ambient
temperature is too high
during the pressure
adjustment.
Take necessary
measures to lower
ambient temperature,
e.g. through venting.
0300A
% RH short-
circuit
Short-circuit in humidity
sensor
Contact Testo Service
0300B
% RH sensor
broken
The humidity sensor is
damaged (sensor
broken)
Contact Testo Service
0300C T
short-circuit
Short-circuit in
temperature sensor
Contact Testo Service
0300D
T sensor broken The temperature sensor
is damaged (sensor
broken)
Contact Testo Service