5 Parameterizing, adjusting and analyzing software (P2A software)
67
Field
Explanation
°C
td
/°F
td
Selection of the unit; only for dewpoint
adjustment.
°C/°F
Selection of the unit; only for temperature
adjustment.
Current reading
Reading in the selected unit °C
td
/°F
td
or °C/°F.
Readings are updated every second.
Reference value
Entry of the read-off value from the reference
measuring instrument.
Permissible entries:
•
max. 10 °C
td
deviation (sum of all 1-point
adjustments)
•
max. 2 K (°C) deviation (sum of all 1-point
adjustments)
> Click
on
[Set offset to zero]
to reset a transmitted reference
value.
-
The current reading is used again.
6.
Click on
[OK]
or
[Cancel]
to close the dialogue.