7 AK Host Operation
34
ANOR
read the currently set standard conditions: standard temperature and
pressure
command: ANOR K0
response:
ANOR E TTT PPP
E: error status 0...9
TTT: temperature [K]
format: floating point example: '273.15'
PPP: pressure [mbar]
format: floating point example: '1013.25'
ATD1
reads PND1 temperature
command: ATD1 K0
response:
ATD1 E TTT
E: error status 0...9
TTT: temperature [°C] for ECE R83: 150°C
format: floating point example: '150.8'
AVD1
read setpoint and actual value of primary dilution factor DF
PND1
if PND1 rotating disk is stopped and therefore DF
PND1
is infinite, response is
'Inf'
command: AVD1 K0
response:
AVD1 E DDD RRR
E: error status 0...9
DDD: DF
PND1
primary dilution factor setpoint
15...300 or 150...3000, depenting on used disk
format: floating point example: '20.0'
RRR: DF
PND1
primary dilution factor actual
value
format: floating point example: '20.2'
AVD2
read setpoint and actual value of secondary dilution factor DF
PND2
command: AVD2 K0
response:
AVD2 E DDD RRR
E: error status 0...9
DDD: DF
PND2
setpoint range: 1...11
format: floating point example: '10.0'
RRR: actual DF
PND2
format: floating point example: '10.05'
ATET
read evaporation tube ET temperature
command: ATET K0
response:
ATET E TTT
E: error status 0...9
TTT: temperature [°C] range: 20...400°C ECE
R83: 300...400°C
format: floating point example: '300.5'
AFD2
read the secondary dilution air supply flow Q
AS
in PND2
command: AFD2 K0
response:
AFD2 E FFF
E: error status 0...9
FFF: flow in lN/min range: 0...15.0 lN/min
format: floating point example: '3.0'
7.5.4.
AK Errors List
There are a number of errors which can be detected by the system and read out using the 'ASTF'
command. These error codes are listed below: