trigtype
Results
0
Immediate trigger
Takes data immediately after GET command
1
Manual trigger
Takes data when START/STOP is pressed
2
Rising edge/rising edge
Takes data after input crosses threshold voltage
3
Falling edge/falling edge
Takes data after input crosses threshold voltage
4
Rising edge/falling edge
Takes data after input crosses threshold voltage
5
Falling edge/rising edge
Takes data after input crosses threshold voltage
6
Single sample
Takes one data point each time
START/STOP is pressed
trigchan
Results
0
Disables the trigger
1
Hardware or software
(hardware trigger only for
Command 1 operation 5, 6, 7;
software trigger for all others)
Triggers on channel 1; channel must be active
2
Software only
Triggers on channel 2; channel must be active
3
Software only
Triggers on channel 3; channel must be active
11
Software only
Triggers on channel 11; channel must be
active
trigthresh
Results
-
channel limit to + channel limit
Begins taking data when signal crosses
threshold in trigtype direction
[channel limit is determined by the sensor that is attached to the channel]
1V
default
prestore
Results
0
% to 100%
Retains this much data from before triggering
(extclock)
This parameter is ignored
rectime
Results
0
None
Does not record time during sampling
1
Absolute
Records absolute time
2
Relative
Records relative time
Note:
This default is different from the original CBL™. Default on original CBL was 0.
filter
Results
0
No filtering
Disables the filtering process (RT* and
NON-RT**)
1
Uses Savitzsky-Golay 5 point filter (NON-RT)
2
Uses Savitzsky-Golay 9 point filter (NON-RT)
3
Uses Savitzsky-Golay 17 point filter (NON-RT)