TMP006EVM Hardware Overview
2.4
Signal Definition of H2 (10-Pin FFC Connector)
shows the signals connected to the H2 connector on the TMP006 Test Board.
Table 4. Signal Definition for H2 (10-Pin FFC
Connector) on TMP006EVM Board
Pin No.
Signal
1
SCL
2
VCC
3
SDA
4
VCC
5
SDO
6
GND
7
SCLK
8
GND
9
CS
10
SDI
3
TMP006EVM Hardware Overview
If not already assembled, the basic hardware setup for the TMP006EVM involves connecting the TMP006
Test Board to the SM-USB-DIG and then connecting the USB cable. This section presents the details of
this procedure.
3.1
Electrostatic Discharge Warning
CAUTION
Many of the components on the TMP006EVM are susceptible to damage by
electrostatic discharge (ESD). Customers are advised to observe proper ESD
handling precautions when unpacking and handling the EVM, including the use
of a grounded wrist strap at an approved ESD workstation.
3.2
Typical TMP006EVM Hardware Setup
Connect the right-angle female socket (H1) on the TMP006 Test Board to the right-angle male header
(H2) on the SM-USB-DIG. Take special care to ensure that the two 10-pin sockets directly align with each
other. Plug the female USB-A cable to the SM-USB-DIG and then plug the male USB-A cable into the
computer.
Always connect the two boards together before connecting the USB cable to avoid any issues if the
connectors are misaligned.
Figure 5. Typical Hardware Connection
7
SBOU109A
–
May 2011
–
Revised October 2011
TMP006EVM User Guide and Software Tutorial
Copyright
©
2011, Texas Instruments Incorporated