Basic Test Procedure with TSW1400
22
SLAU374B – December 2011 – Revised May 2016
Copyright © 2011–2016, Texas Instruments Incorporated
TSW308x Evaluation Module
Figure 17. Load File to Transfer into TSW1400
4. For the TSW3084, select the file "WCDMA_TM1_complexIF30MHz_Fdata307.2MHz_1000.tsw" under
C:\Program Files\Texas Instruments\High Speed Data Converter Pro\1400 Details\Testfiles
5. For the TSW30H84 and TSW3085, select the file
"WCDMA_TM1_complexIF30MHz_Fdata614.4MHz_1000.tsw" under C:\Program Files\Texas
Instruments\High Speed Data Converter Pro\1400 Details\Testfiles
6. Enter 307.2M or 614.4M for the "Data Rate" and 2's complement for the "DAC Option".
7. Select Hanning for "Window".
8. In the “DAC Selection” panel on the left side of the GUI, click on “Send” to load the data into memory.
9.
Toggle the SIF SYNC button of the TSW308x EVM GUI to synchronize the appropriate digital
blocks, if the example file with NCO setting is used.
10. Verify the spectrum using the spectrum analyzer at the RF output port(s), J7 or J9.
11. With 1780 MHz of LO, the expect results are shown in
(TRF3705 Low-Gain Mode) and
(TRF3705 High-Gain Mode).